Support_FAQ Banner

FAQs

  • How to do when the radio of solvents is not accurate?

    Clean the solvent filter head completely to remove any impurities, It is best to use ultrasonic cleaning.

  • What cause the high baseline noise ?

    1. flow cell of the detector was polluted.

    2. Low energy of light source.

    3. Influence of pump pulse.

    4. Temperature effect of detector.

    5. There are bubbles in the test pool.

    6. Column or mobile phase contamination.

    In preparative chromatography, a small amount of baseline noise has little effect on separation.

  • How to do if liquid level alarm abnormally?

    1. The tube connector at the back of the machine is loose or damaged; Replace the tube connector;

    2. Gas way check valve is damaged. Replace the check valve.

  • How to do if the historical record prompts

    After the separation, it is necessary to wait 3-5 minutes before shutting down to ensure the integrity of the experiment records.

  • Why do we need to equilibrate the column before separation?

    Column equilibration can protect the column from being damaged by exothermic effect when solvent quickly flushing through the column. While dry silica pre-packed in the column being contacted by the solvent for first time during separation run, a lot of heat might be released especially when the solvent flushes in a high flow rate. This heat might cause the column body to deform and thus solvent leakage from the column. In some cases, this heat might also damage heat sensitive sample.

  • How to do when the pump sound louder than before?

    It maybe caused by the lack of lubricating oil at the rotating shaft of the pump.

  • What is the volume of tubings and connections inside the instrument?

    The total volume of system tubing, connetors and mixing chamber is about 25 mL.

  • How to do when negative signal responce in the flash chromatogram, or the eluting peak in the flash chromatogram is abnormal…

    The flow cell of the detector module is contaminated by the sample which has strong UV absorption. Or it might be due to solvent UV absorption which is a normal phenomenon. Please do the following operation:

    1. Remove the flash column and flush the system tubing with strongly polar solvent then followed by weakly polar solvent.

    2. Solvent UV absorption problem: e.g. while n-hexane and dichloromethane (DCM) are employed as the eluting solvent, as the proportion of DCM increases, the baseline of chromatogram may continue to be below zero on Y-axis since the absorption of DCM at 254 nm is lower than that of n-hexane. In case of this phenomenon happens, we can handle it by clicking the “Zero” button on the separation running page in SepaBean App.

    3.The flow cell of the detector module is heavily contaminated and needs to be cleaned ultrasonically.

  • How to do when the column holder head does not lift up automatically ?

    It might be due to that the connectors on the column holder head as well as on the base part are swelled by solvent so that the connectors are stuck.

    User can manually lift up the column holder head by using a little bit force. When the column holder head is lifted up to a certain height, the column holder head should be able to be moved by touching the buttons on it. If the column holder head cannot be lifted up manually, user should contact the local technical support.

    Emergency alternative method: User can install the column on the top of the column holder head instead. Liquid sample can be injected directly onto the column. Solid sample loading column can be installed on the top of the separation column.

  • How to do if the intensity of detector become weak ?

    1. Low energy of light source;

    2. The circulation pool is polluted; Intuitively, there is no spectral peak or the spectral peak is small in the separation , The energy spectra shows a value of less than 25%.

    Please flush the tube with an appropriate solvent at 10ml/min for 30min and observe the energy spectrum.If there is no change in the spectrum,it seems low energy of light source,please replace the deuterium lamp; If the spectrum changed, the circulation pool is polluted,please continue to clean with appropriate solvent.

  • How to do when the machine leaks fluid inside ?

    Please check the tube and connector regularly.

  • How to do if baseline keep drifting upward when ethyl acetate was employed as the eluting solvent?

    The detection wavelength is set at the wavlength lower than 245 nm since ethyl acetate has strong absorption at the detection range lower than 245nm. The baseline drifting will be most dominant when ethyl acetate is used as eluting solvent and we choose 220 nm as the detection wavelength.

    Please change the detection wavelength. It is recommended to choose 254nm as the detection wavelength. If 220 nm is the only wavelength suitable for the sample detection, user should collect the eluent with carefully judgement and excessive solvent might be collected in this case.