Power on the instrument and wait for its prompt "Ready". Make sure the iPad network connection is correct, and the router is powered on.
Check and confirm router status to make sure the iPad can be connected to the current router.
The equilibration is done when the column is totally wetted and looks translucent. Usually this can be done in flushing 2 ~ 3 CVs of the mobile phase. During the equilibration process, occasionally we might find that the column cannot be completely wetted. This is a normal phenomenon and will not compromise the separation performance.
Check if the tube rack is correctly placed in the right position. When this is done, the LCD screen on the tube rack should show a connected symbol.
If the tube rack is faulty, user can choose customized tube rack from the tube rack list in SePaBean App for temporary use . Or contact after-sale engineer.
Check whether the solvent bottle is lack of related solvent and replenish the solvent.
If the solvent line is full of solvent, please do not worry. Air bubble does not affect the flash separation since it is inevitable during solid sample loading. These bubbles will be gradually drained out during separation procedure.
Please open the back cover of the instrument, clean the pump piston rod with ethanol (analysis of pure or above), and rotate the piston while washing until the piston turns smoothly.
1. Instrument will not be able to pump the solvents when ambient temperature above 30℃, especially low boiling solvents, Such as dichloromethane or Ether.
Please ensure that the ambient temperature is below 30℃.
2. Air occupy the pipeline while the instrumnet out of operation for long time.
Please add ethanol to the ceramic rod of the pump head (analysis of pure or above) and increase the flow rate at the same time. The connector in front of the pump damaged or Loose, this will cause the line to leak air .Please check carefully whether the pipe connection is loose.
3. The connector in front of the pump damaged or Loose, it will cause the line to leak air .
Please confirm whether the pipe connector is in good condition.
The collect valve are blocked or aging. Please replace the three-way solenoid valve.
ADVICE: Please contact the after-sale engineer to deal with it.
Clean the solvent filter head completely to remove any impurities, It is best to use ultrasonic cleaning.
1. flow cell of the detector was polluted.
2. Low energy of light source.
3. Influence of pump pulse.
4. Temperature effect of detector.
5. There are bubbles in the test pool.
6. Column or mobile phase contamination.
In preparative chromatography, a small amount of baseline noise has little effect on separation.
1. The tube connector at the back of the machine is loose or damaged; Replace the tube connector;
2. Gas way check valve is damaged. Replace the check valve.
After the separation, it is necessary to wait 3-5 minutes before shutting down to ensure the integrity of the experiment records.
Column equilibration can protect the column from being damaged by exothermic effect when solvent quickly flushing through the column. While dry silica pre-packed in the column being contacted by the solvent for first time during separation run, a lot of heat might be released especially when the solvent flushes in a high flow rate. This heat might cause the column body to deform and thus solvent leakage from the column. In some cases, this heat might also damage heat sensitive sample.
It maybe caused by the lack of lubricating oil at the rotating shaft of the pump.
The total volume of system tubing, connetors and mixing chamber is about 25 mL.
The flow cell of the detector module is contaminated by the sample which has strong UV absorption. Or it might be due to solvent UV absorption which is a normal phenomenon. Please do the following operation:
1. Remove the flash column and flush the system tubing with strongly polar solvent then followed by weakly polar solvent.
2. Solvent UV absorption problem: e.g. while n-hexane and dichloromethane (DCM) are employed as the eluting solvent, as the proportion of DCM increases, the baseline of chromatogram may continue to be below zero on Y-axis since the absorption of DCM at 254 nm is lower than that of n-hexane. In case of this phenomenon happens, we can handle it by clicking the “Zero” button on the separation running page in SepaBean App.
3.The flow cell of the detector module is heavily contaminated and needs to be cleaned ultrasonically.
It might be due to that the connectors on the column holder head as well as on the base part are swelled by solvent so that the connectors are stuck.
User can manually lift up the column holder head by using a little bit force. When the column holder head is lifted up to a certain height, the column holder head should be able to be moved by touching the buttons on it. If the column holder head cannot be lifted up manually, user should contact the local technical support.
Emergency alternative method: User can install the column on the top of the column holder head instead. Liquid sample can be injected directly onto the column. Solid sample loading column can be installed on the top of the separation column.
1. Low energy of light source;
2. The circulation pool is polluted; Intuitively, there is no spectral peak or the spectral peak is small in the separation , The energy spectra shows a value of less than 25%.
Please flush the tube with an appropriate solvent at 10ml/min for 30min and observe the energy spectrum.If there is no change in the spectrum,it seems low energy of light source,please replace the deuterium lamp; If the spectrum changed, the circulation pool is polluted,please continue to clean with appropriate solvent.
Please check the tube and connector regularly.
The detection wavelength is set at the wavlength lower than 245 nm since ethyl acetate has strong absorption at the detection range lower than 245nm. The baseline drifting will be most dominant when ethyl acetate is used as eluting solvent and we choose 220 nm as the detection wavelength.
Please change the detection wavelength. It is recommended to choose 254nm as the detection wavelength. If 220 nm is the only wavelength suitable for the sample detection, user should collect the eluent with carefully judgement and excessive solvent might be collected in this case.
Clean the solvent filter head completely to remove any impurities. Use ethanol or isopropanol to flush the system completely to avoid immiscible solvent problems.
To clean the solvent filter head, disassemble the filter from the filter head and clean it with a small brush. Then wash the filter with ethanol and blow-dry it. Re-assemble the filter head for future use.
Either switch from normal phase separation to reversed phase separation or vice versa, ethanol or isopropanol should be used as the transition solvent to completely flush out any immiscible solvents in the tubing.
It is suggested to set the flow rate at 40 mL/min to flush the solvent lines and all the internal tubings.
Please reposition the bottom of the column holder after Loosen the screw.
1. The system flow rate is too high for the current flash column.
2. Sample has poor solubility and precipitates from the mobile phase, thus resulting tubing blockage.
3. Other reason causes tubing blockage.
The environment is too wet, or the solvent leakage to the inside of the column holder causes short circuit. Please heat the column holder properly by a hair dryer or a hot air gun after power off.
Solvent leakage might be due to the solvent level in the waste bottle is higher than the height of the connector at the base of the column holder.
Place the waste bottle below the operation platform of the instrument, or quickly move down the column holder after removing the column.
This cleaning function is designed to clean the system pipeline before separation run. If "post-cleaning" has been performed after the last separation run, this step could be skipped. If it is not performed, it is recommended to do this cleaning step as instructed by the system prompt.